Analysis of THC in Whole Blood
来源: | 作者:DPX | 发布时间: 1986天前 | 6284 次浏览 | 分享到:

Analysis of THC in Whole Blood



Automation Compatibility: Hamilton, Integra

Cannabis is the most widely abused illicit drug in the US. Results from the 2013 National Survey on Drug Use and Health show that between 2007 and 2012 the rate of marijuana use rose from 5.8% to 7.3%. As use of marijuana increases, it becomes more prevalent in clinical and forensic case work, particularly in impaired driving cases. In 2007 the National Highway Traffic Safety Administration (NHTSA) reported that 8.6% of nighttime drivers tested positive for cannabinoids—a rate almost four times higher than those with blood alcohol levels equal to or above 0.8 g/L. Recent marijuana use is associated with a 2–6 fold increased crash risk, depending on dose, than when unimpaired.

The LC-MS/MS method developed in this study minimizes required sample volume and provides sensitive quantitation of THC, 11-OH-THC and THC-COOH in whole blood. Notably, our method simplifies sample preparation for analysis of THC and its metabolites in whole blood using an automated dispersive pipette extraction without subsequent dilution or solvent evaporation. The extraction process is rapid, minimizes matrix effects (~ 30% or less), and maximizes analyte recoveries (>93%). LODs and LOQs were below 0.75 ng/ mL and 2 ng/mL, respectively. These outcomes clearly provide the necessary sensitivity to meet laboratory cut-offs with minimal imprecision (<8%). All calibrations were linear (R2 >0.99) over two orders of magnitude (0.5–50 ng/mL).

This work was further expanded upon by the National Institute of Drug Abuse (NIDA) to include additional cannabinoid metabolites and application to urine.


1

Blood + I.S. + ACNN

Vortex & Centrifuge (for increased efficiency use Tip-on-Tip)

2

Transfer supernatant to well platewith 500 μL acetonitrile

3

Load Nimbus96 Deck – Click Play

4

Add 0.1 M FA to Sample

Aspirate & Dispense 4X with XTR tips with WAX-S sorbent

Transfer Supernatant

5

Inject onto LC-MS/MS


Publications Referencing this method:

  1. Quantification of cannabinoids and their free and glucuronide metabolites in whole blood by disposable pipette extraction and liquid chromatography-tandem mass spectrometry; Journal of Chromatography A
  2. Simultaneous quantification of 11 cannabinoids and metabolites in human urine by liquid chromatography tandem mass spectrometry using WAX-S tips; Journal of Analytical and Bioanalytical Chemistry 
  3. Quantitation of Cannabinoids in Breath Samples Using a Novel Derivatization LC–MS/MS Assay with Ultra-High Sensitivity; Journal of Analytical Toxicology 

  • 三种 Protein A树脂纯化效率比较及流程优化

    2021-09-20

    简要介绍 dSPE 在全自动移液工作站上应用实现全自动抗体纯化的原理。
    比较了三种 Protein A 纯化 HIgG 的效率及纯化流程优化方法。

  • 使用 IMCSTips 实验全自动纯化(一步)方法及数据分析

    2021-09-20

    1. IMCStips 应用原理展示
    2. His-tagged 蛋白纯化流程优化及数据分析

  • 全自动抗体、蛋白纯化演示

    2021-09-20

    IMCStips 亲和层析柱
    ● 一次层析即可快速产生高纯重组蛋白
    ● 一致性好、回收率高,个样品的差异不到 10%
    ● 可结合各种复杂方法的模板化工作流
    ● 与 Hamilton 移液工作站结合,可在 30 分钟内处理 96 个样品
    ● 可定制的工作流程

  • 振荡器-BenchWaver

    2020-02-08

    BenchWaver™ 把传统的平台摇摆运动与轨道振动器的圆周运动结合起来,这种最优组合能够产生最适合分子生物学应用的三维运动。

  • 均质机-D1030 演示

    2020-02-08

    强力混匀,适于细胞裂解、研磨或均质
    与通用的组织研磨机更快速、更高效
    采用密封2.0mL管,消除交叉感染
    中通量,可同时处理3~6个样品